Data Sheet User's Manual Calibration Procedure Application Notes
DIFFERENTIAL PROBE SI-200
Calibration Procedure

This procedure is for use by qualified service personnel to adjust SI-200 properly. The equipments required are listed in Table 1
No.
Item
Minimum Requirements
1
Power Supply 6VDC/90mA or 9VDC/70mA  mains adaptor or 4xAA cells or power lead.
2
DMM DC Accuracy &lt 0.5%
3
Function Generator Maximum Frequency >10MHz
Sine-wave Distortion <1%
4
Oscilloscope Bandwidth >= 20MHz
Accuracy <= 2%
Table 1
 
This procedure is divided into following steps;
A. Prepare the probe for adjustment.
B. Adjust input bias voltage.
C. Adjust output offset voltage.
D. Adjust CMRR at 60Hz.
E. Adjust square-wave compensation for +input.
F. Adjust square-wave compensation for -input.
G. Adjust CMRR at 10MHz.
A. Prepare the probe for adjustment.
ˇ@A-1. Use a small flat screwdriver to peel the both panels off, referring to figure 1.
 

Fig. 1

ˇ@A-2. Loosen the four screws on the plastic cases.
ˇ@A-3. Remove the plastic cases. Figure 2 shows the location of adjustments on bottom side of the metal case.
ˇ@A-4. Connect the power source.
ˇ@A-5. Allow the probe and test equipments to warm up 20 minutes at an ambient temperature of 18 degree Celsius to 28 degree Celsius.
ˇ@A-6. In order to make following operation easier, use one plastic case to support the metal case and the input head.
 

Fig. 2

B. Adjust input bias voltage
ˇ@B-1. Connect the probe as shown in figure 3.
ˇ@B-2. The value of input bias voltage now is displayed on the digital multimeter.
ˇ@B-3. Adjust VR2 to make the input bias voltage as small as possible.
ˇ@B-4. The criterion is -2mV <= Vib <= +2mV.
 

Fig. 3

C. Adjust output offset voltage
ˇ@C-1. Connect the probe as shown in figure 4.
ˇ@C-2. Adjust VR3 for minimum output offset voltage.
ˇ@C-3. The criterion is -5mV <= Vout <= +5mV.
 

Fig. 4

D. Adjust CMRR at 60Hz
ˇ@D-1. Connect the probe as shown in figure 5.
ˇ@D-2. Set the output of the function generator to 20Vp-p 60Hz sine-wave, and set the input impedance of the oscilloscope to 50ohm. (Add a feedthrough 50ohm terminator to the input, if the oscilloscope doesnˇ¦t provide the 50ohm input impedance.)
ˇ@D-3. Adjust VR1 for minimum Vout displayed on the oscilloscope.
ˇ@D-4. The criterion is Vout <= 0.2mVp-p.


Fig. 5

E. Adjust square-wave compensation for +input.
ˇ@E-1. Connect the probe as shown in figure 6.
ˇ@E-2. Set the output of the function generator to 1.4Vp-p, 100KHz square-wave.
ˇ@E-3. Set the input impedance of the oscilloscope to 50ohm (Add a feedthrough 50ohm terminator to the input, if the oscilloscope doesnˇ¦t provide the 50ohm input impedance.)
ˇ@E-4. Adjust VC1 to make the front corner roll off or overshoot of the square-wave displayed on the oscilloscope less than 2.8mV.
 

Fig. 6

F. Adjust square-wave compensation for -input.
ˇ@ F-1. Connect the probe as shown in figure 7.
ˇ@F-2. Set the output of the function generator to 1.4Vp-p, 100KHz square-wave.
ˇ@F-3. Set the input impedance of the oscilloscope to 50ohm (Add a feedthrough 50ohm terminator to the input, if the oscilloscope doesnˇ¦t provide the 50ohm input impedance.)
ˇ@F-4. Adjust VC2 to make the front corner roll off or overshoot of the square-wave displayed on the oscilloscope less than 2.8mV.
 

Fig. 7

G. Adjust CMRR at 10MHz
ˇ@ G-1. Connect the probe as shown in figure 8.
ˇ@G-2. Set the output of the function generator to be 20Vp-p, 10MHz, sine-wave, and set the input impedance of the oscilloscope to be 50ohm (Add a feedthrough 50ohm terminator to the input , if the oscilloscope doesnˇ¦t provide the 50ohm input impedance.)
ˇ@G-3. Adjust VC1 slightly for minimum Vout displayed on the oscilloscope.
ˇ@G-4. The criterion is Vout <= 6mVp-p.
 

Fig. 8


   Sapphire Instruments Co.,Ltd.ˇ@ˇ@ˇ@ˇ@ˇ@

16F, 866-9, Jung-Jeng Rd., Jung-Ho City, 235 Taipei County, Taiwan, ROC

TEL: 886-2-22250572 (Rep.)     FAX: 886-2-22253412

Email : tina@sapphire.com.tw     Web Site : www.sapphire.com.tw

ˇ@