|
|
DIFFERENTIAL PROBE SI-51
Calibration Procedure
|
|
|
This procedure is for use by qualified service personnel to adjust SI-51 properly.
The equipments required are listed in Table 1.
|
|
No
|
Item
|
Minimum Requirements
|
|
1
|
Power Supply
|
6VDC/90mA or 9VDC/330mA mains adaptor or 4xAA cells or power lead
|
|
2
|
DMM
|
DC Accuracy < 0.5%
|
|
3
|
Function Generator
|
Maximum Frequency > 10MHz
Sine-wave Distortion < 1%
|
|
4
|
Oscilloscope
|
Bandwidth >= 20MHz
Accuracy <= 2%
|
Table 1
|
This procedure is divided into following steps:
A. Prepare the probe for adjustment.
B. Adjust input bias voltage.
C. Adjust output offset voltage.
D. Adjust CMRR at 60Hz.
E. Adjust square-wave compensation for +input.
F. Adjust square-wave compensation for -input.
G. Adjust CMRR at 1MHz. |
|
A. Prepare the probe for adjustment.
|
|
A-1. Use a small flat screwdriver to peel the both panels off, referring to figure 1.
|

Fig. 1
|
|
|
A-2. Loosen the four screws on the plastic cases.
A-3. Remove the plastic cases. Figure 2 shows the location of adjustments on topside and bottom side of the metal case.
A-4. Connect the power source.
A-5. Allow the probe and test equipments to warm up 20 minutes at an ambient temperature of 18 degree Celsius to 28 degree Celsius.
A-6. In order to make following operation easier, use one plastic case to support the metal case and the input head.
|

Fig. 2
|
|
B. Adjust input bias voltage |
B-1. Connect the probe as shown in figure 3.
B-2. The value of input bias voltage now is displayed on the digital multimeter.
B-3. Adjust VRA to make the input bias voltage as small as possible.
B-4. The criterion is -3mV <= Vib <= +3mV.
|
|
|

Fig. 3
|
|
C. Adjust output offset Voltage |
C-1. Connect the probe as shown in figure 4.
C-2. Adjust VRC for minimum output offset voltage.
C-3. The criterion is -1mV <= Vout <= +1mV.
|

Fig. 4
|
|
D. Adjust CMRR at 60Hz
|
|
D-1. Connect the probe as shown in figure 5.
|
|
D-2. Set the output of the function generator to 620Vp-p 60Hz sine-wave, and set the input impedance of the oscilloscope to 50ohm. (Add a feedthrough 50ohm terminator to the input, if the oscilloscope doesnˇ¦t provide the 50ohm input impedance.) |
|
D-3. Adjust VR1 for minimum Vout displayed on the oscilloscope.
|
D-4. The criterion is Vout <= 0.6mVp-p.
|

Fig. 5
|
|
E. Adjust square-wave compensation for +input.
|
|
E-1. Connect the probe as shown in figure 6.
|
|
E-2. Set the output of the function generator to 40Vp-p, 1KHz square-wave.
|
|
E-3. Set the input impedance of the oscilloscope to 50ohm (Add a feedthrough 50ohm terminator to the input, if the oscilloscope doesnˇ¦t provide the 50ohm input impedance.)
|
|
E-4. Adjust VCA to make the front corner roll off or overshoot of the square-wave displayed on the oscilloscope less than 4mV |
|
|

Fig. 6
|
|
F. Adjust square-wave compensation for -input.
|
|
F-1. Connect the probe as shown in figure 7.. Referring to Fig. 7.
|
|
F-2. Set the output of the function
generator to 40Vp-p, 1KHz square-wave. |
|
F-3. Set the input impedance of
the oscilloscope to 50ohm (Add a feedthrough 50ohm terminator to the input, if
the oscilloscope doesn't provide the 50ohm input impedance.)
|
|
F-4. Adjust VCB to make the front
corner roll off or overshoot of the square-wave displayed on the oscilloscope
less than 4mV. |
|
|

Fig. 7
|
|
G. Adjust CMRR at 1MHz
|
|
G-1. Connect the probe as shown in figure 8.
|
|
G-2. Set the output of the function generator to be 20Vp-p, 1MHz, sine-wave, and set the input impedance of the oscilloscope to be 50ohm (Add a feedthrough 50ohm terminator to the input , if the oscilloscope doesnˇ¦t provide the 50ohm input impedance.)
|
|
G-3. Adjust VC1 slightly for minimum Vout displayed on the oscilloscope
|
|
G-4. The criterion is Vout <=
0.7mVp-p.
|
|
|

Fig. 8
|
|
|
   Sapphire Instruments
Co.,Ltd.

16F, 866-9, Jung-Jeng Rd., Jung-Ho City, 235 Taipei County, Taiwan, ROC
TEL: 886-2-22250572 (Rep.) FAX: 886-2-22253412
Email : tina@sapphire.com.tw     Web Site : www.sapphire.com.tw
|